The AcuQor family operates over a universal input range of 85-264V and 47-63Hz making it suitable for World-Wide deployment. Full output power is available up to 50degC with natural convection (derates linearly to 300W at 70degC). A transient power rating of 500W is possible for up to 5 seconds in the open frame version, and even longer in a proprietary encased version. Active Power Factor Correction is incorporated to a level >0.98 enabling compliance with IEC/EN61000-3-2. Both AC lines are internally fused.
The ACuQor models include versions designed for both BF (direct patient contact) and CF (direct cardiac contact) applications in accordance with UL/EN60601-1. Input Earth Leakage Current and type CF Patient Leakage Current are well below the requirements of the standard at <110uA and <4uA respectively. There is also the option of ordering defibrillation rated models that incorporate the minimum output to protective earth creepage & clearance distances and withstand the defibrillator pulse test guidelines of UL/EN60601-1. Conducted noise emissions are below Level B limits of EN55011/55022, FCC part15. A host of additional control and signalling inputs & outputs are offered including a 5V “always on” standby power output, a DC power good signal, AC power good signal, remote enable input, and a fan status output / fan enable input function. Over current, over voltage and over temperature protection with auto recovery is also built into the unit. The operating temperature range is 0 to +70degC.