With built-in active PFC function PF>0.93, these units comply with the harmonic current requirement of EN61000-3-2 and comply to UL60601-1 & TUV EN60601-1 medical safety standards. 100~110W of power can be provided using only free air convection, while the full rated output of 145~160W can be extracted by providing the power supplies 20.5CFM of forced air. Featuring 300 μA of leakage current at 264VAC, they can be equipped into non-patient contact medical equipment and still comply with Class B per EMI regulations. Standard functions include PF/PG signal output, remote sense (single output 5~15V), and short circuit, overload, over voltage, and over temperature protections. General medical usage that requires low leakage current, general electronic application, and 1U low profile IT equipment are all suitable for applying this miniature product series.
RPSG/RPDG/RPTG-160 are specially designed with additional PS-ON control and 5V/0.8A standby output, these green models can achieve 0.75W of no-load-power-consumption and are suitable for use in final systems regulated by Energy Star/EuP norms or any kind of energy-saving system design.