Certified to both medical and industrial safety standards, typical applications include home healthcare, hospital equipment, imaging and clinical diagnostic systems in addition to industrial, test and measurement and communications equipment. With a low 37mm profile the XMS500 can be easily incorporated into limited space environments.
The flexibility of XMS500 allows engineers to select from a wide range of options to optimise system design and cost. These include standby voltages, fan supply voltage, case options (including built-in, low audible noise, fan), remote on/off, AC fail and single or dual input fusing. Standard output voltages offered are 12, 24, 36 and 48V. Additional voltages and options are available.
The product series will operate from an input of 90-264Vac, and is able to withstand 300Vac for five seconds. The XMS500 will operate in ambient temperatures of up to 70°C, derating linearly above 50°C by 2.5% per °C. The open frame models measure 102x180x37mm and the U channel 107x180x39.5mm.
All models are safety certified to IEC/EN/ES 60601-1, IEC/EN/UL 60950-1 with CE marking for the Low Voltage and RoHS2 Directives. The XMS500 also complies with EN 55011 level B and EN 55032 level B conducted and radiated emissions (for both the Class I and II products), and meets the EN 61000-3-2 harmonics, IEC60601-1-2 Edition 4 and IEC 61000-4 immunity standards.
Input to output isolation is 4kVac (2 x MOPP), input to ground isolation 1.5kVac (1 x MOPP) and output to ground isolation is 1.5kVac (1 x MOPP) making it suitable for both B and BF rated equipment.