The AcuQor family operates over a universal input range of 85-264Vrms and 47-63Hz making it suitable for World-Wide deployment. Through the use of SynQor’s “QorCool” thermal conduction techniques, full output power is available when attached to a cold plate maintained at 50oC. It has a transient power rating of 700W for up to 15 seconds. Active Power Factor Correction is incorporated to a level >0.98 enabling compliance with IEC/EN61000-3-2. Both Line and Neutral are internally fused.
The AQ0500M series includes versions designed for both BF (direct patient contact) and CF (direct cardiac contact) applications in accordance with UL/EN60601-1. Input Earth Leakage Current and type BF/CF Patient Leakage Current are well below the requirements of the standard at <125uA and 2uA respectively. There is also the option of ordering defibrillation rated models that incorporate the minimum output to protective earth creepage & clearance distances and withstand the defibrillator pulse test guidelines of UL/EN60601-1. Conducted noise emissions are below Level B limits of EN55011/55022, FCC part 15. A 5V “always on” standby power output, DC power good signal, AC power good signal, remote enable input, and a fan status output function are standard. Over current, over voltage and over temperature protection with auto recovery is also built into the unit. The operating temperature range is 0 to +70oC. An industrial grade version is also available for those applications where size, weight, efficiency and reliability are essential.