Both of the series operate with an input voltage range of 85 to 264VAC and deliver 5, 12, 15, 24 and 48VDC output voltages, which can be trimmed over a ±10% range to meet the exact requirements of the application. All models feature efficiency of up to 93.5%, require no minimum load and can be used at altitudes of up to 5000m at temperatures between -40 and +85°C.
The output voltages are fully regulated and have tolerances of less than ±0.2% over the entire input voltage range and less than ±0.5% over the entire load range. These space-saving 3×2” modules offer 4kVAC reinforced isolation between input and output, 2.5kVAC between output and case and also meet the requirements for medical applications with patient contact (2xMOPP at 250VAC working voltage, 8mm creepage & clearance).
These modules are approved according to medical safety standard IEC/ES/EN-60601-1 3rd Edition and feature BF-rated outputs with less than 75µA leakage current. All models have a built-in Class B EMI filter and come with five year warranty.